Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Operating Instructions E-Drive
134°C
62 61 957 D3436.201.05.05.
7.5

Sterilizing

In accordance with ISO 17664/RKI guidelines, the contra-angle
handpiece is classified as critical B (sterile).
✔ The instrument is clean and disinfected.
274°F
✔ Instrument maintenance is complete.
✔ The instrument can be sterilized in packaging suitable for
sterilization and storage: paper/plastic composite packaging
according to ISO 11607.
➢ Sterilize the instrument in the steam sterilizer with saturated water
vapor.
Overpressure:
Temperature:
Holding time:
2.04 bar (29.59 psi)
134°C (274° F)
3 min.
7 Conditioning
45

Publicidad

Tabla de contenido
loading

Productos relacionados para Komet E-Drive

Tabla de contenido