Biotech Mode - Covidien Genius 2 Manual De Instrucciones

Ocultar thumbs Ver también para Genius 2:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 78

Biotech Mode

The biotech function is used to select the operational mode of the thermometer and to verify the installed software version. All
operational mode settings in biotech mode are stored in nonvolatile memory and retained through system power cycles, such as
changing batteries. All factory calibration parameters are also stored in nonvolatile memory.
The factory default settings are shown below:
Temperature mode °C (unlocked)
Site mode
Ear
Site text
On
The biotech mode is accessed by pressing and holding the timer and °C/°F buttons for four seconds. All LCD segments will light for
one second, the thermometer will issue a single beep, and the display will show scrolling dashes. Pressing the timer button cycles
through the biotech modes. When options are available within a mode, the °C/°F button cycles through the options.
Pressing the timer button after the site text display will return the user to the installed software version. To exit biotech mode,
two options are available: (1) press and hold the °C/°F and timer buttons for one second, or (2) the device will automatically exit
biotech mode after about 30 seconds of inactivity. Any changes are saved.
The biotech mode sequence is shown below:
Software version
Displays the installed software version of device. Where "00" is the current software version.
Temperature mode
°C (unlocked)
Site mode
Oral
Site text
Pressing the °C/°F button when in this mode turns the body site text labels on or off. The labels will remain on when an "X" appears
inside the box icon, and the text will remain off when the box is empty.
Tympanic Thermometer and Base
Locked °C
Core
Locked °F
Rectal
8
°F (unlocked)
Ear
Table of Contents
en

Publicidad

Tabla de contenido
loading

Tabla de contenido