This equipment should be operated by dental surgeons for oral and maxillo-facial exams and
procedures only.
Do not use the equipment with water pressure, compressed air or electrical voltage outside the
specifications given in chapter 5 (Installation Requirements) of this manual. Using the equipment
outside the specified conditions may cause a loss of functionality. Equipment defects resulting from
the equipment used outside of its specifications will not be covered by warranty.
Do not replace the fuse in the chair base while it is switched on. There's a risk of electric shock!
Turn Off the On/Off switch on the chair base before replacing the fuse.
Do not remove the plastic covers from the equipment. There's a risk of electric shock! Only the
accredited technician is authorized to perform this procedure.
Do not install or use any electrical equipment over or near the Olsen's Dental Unit. If necessary, the
equipment must be checked to see if it is functioning normally in the configuration in which it will
be used.
The patient cannot operate this equipment.
Do not perform maintenance or cleaning procedure of the equipment while the equipment is on or
with the patient.
Do not perform the following procedures if it is possible to touch the patient, even if unintentionally,
during the procedure:
• Coupling or removal of the electrical micromotor from its coupling;
• Coupling or removal of the scaler from its coupling;
• Coupling or removal of high-speed optical fiber handpiece from its coupling;
• Fuse replacement;

10.2 - Disposal

Debris, residues and infectious materials resulting from the procedures performed on this
equipment must be deposited in biological waste duly identified and in accordance with current
legislation.
For proper disposal of this equipment and its components and accessories, we recommend that it
be sent to specialized recycling companies to ensure the best destination of each component
without harm to the environment.

10.3 - Transport and Storage

It is recommended that the equipment transportation and storage be made in its original packaging.
Transport carefully protecting equipment from falls and impacts.
Protect from moisture, rain exposure and direct contact with liquids.
Keep it sheltered from the sun.
Do not heap up more than 4 volumes.
Temperature range for transportation and storage: -10°C to +45º C.
Moisture limits for transport and storage: 20% to 70%.
Lift the chair only by its base.
37