2.1.4 - Movement Interruption

Equipment that features automatic movements such as Work Position and Zero Position has the movement
interrupted immediately with the execution of any movement command of the chair, either on the remote foot
pedal (Logic and Infinity), on the chair base controls (Sprint and Quality) or Blue Touch Panel. When you cancel an
automatic movement, the equipment emits 3 beeps.
If the user sees any situation that may generate any kind of risk to the patient, stop the chair immediately.
The user must supervise all equipment automatic movements.

2.1.5 Movements Blocking (Optional)

While the propulsion/activation pedal is press, all movements of the chair will be blocked so that there is
no accidental movement.

2.1.6 Emergency Stop Button (Optional)

The emergency stop button (1) interrupts the electrical supply of all the equipment and
consequently interrupts any movement of the chair.

2.1.7 Digital Voltmeter (Optional)

The digital voltmeter (1) will show the voltage of the equipment's power supply network
during operation. This device is a standard item for the Logic and Infinity models.

2.1.8 - Anti-crushing System (Optional)

Consists of devices (1) located on the
chair base, which when activated interrupt the
movement "Seat Down", perform the "Seat Up"
movement for 4 seconds and trigger a long
beep (10 seconds), preventing that any object
or part of the body that is on the chair base is
crushed. After the Anti-crush system is
activated, the equipment will not allow the commands to be used
again. To do this, turn it Off (2) and then On again.

2.1.9 - Positioning of the Patient, the User, and Others

The patient should sit on the chair seat, with the legs over the seat, with the back resting on the backrest
and the hands-on the armrests or over his own body. The user must adjust the headrest according to the dental
procedure to be performed.
To ensure safe and proper patient positioning, the user should advise the patient to keep his hands insight
during the entire equipment movement.
The user must stay near the headrest or the sides of the backrest, observing the distance required for
performing the dental procedures. To move the seat and backrest, it is recommended that the user keeps a
minimum distance of 30 cm, avoiding staying on the chair movement line.
Other persons should keep a minimum distance of 50 cm from the equipment during the backrest and the
seat movements.
The presence of the user, persons or objects in the equipment and its components movement areas
may cause damage to the equipment and/or impair its proper working.