Do not move or store the equipment on uneven surfaces.

10.4 - Contraindications

This equipment is contraindicated for any use other than that for which it is intended or to be operated by
unqualified personnel.

10.4.1 - Curing Light Contraindications

The curing light use in cardiac, pregnant or child patients should be performed with precautions.

10.4.2 - Contraindications on the Scaler Use

It's forbidden the scaler use of hemophilic patients.
Patients, dentists or dental assistants with pacemakers are prohibited from using or approaching the scaler
during its use.
The scaler uses in cardiac, pregnant or child patients should be performed with precautions.
For solving possible problems in a simple and practical way, just follow the instructions in the following
tables:
Item
Problem
The chair doesn't perform
1
any command
The chair doesn't memorize
2
work positions
The working table does not
3
lock in position
4
Operating light is not lighting
Micromotor/turbine doesn't
5
work or is weak
Pneumatic handpiece has
6
water leakage on the
coupling
The pneumatic handpiece
7
doesn't have water on the
spray
The saliva ejector is weak or
8
loses suction during the
procedure
9
The curing light isn't working

11 - Troubleshooting

Causes
1º-The equipment is not connected to the mains
electricity
2º-Electrical circuit breaker is switched off
3º-There is no power on the mains electricity
4°-Emergency Stop button is activated
5º-The protection fuse is blown
1º-The equipment is not performing the movement
2º-Is the record position command incorrect
3º-Is there an electronic problem
1º-The air compressor is not working properly
2º-Pneumatic problem or on the capacitive sensor
prbem
1º-The equipment does not perform the movements
2º-The LED is burned
3º-A blown fuse or component rupture
1º-The handpiece ducts are clogged
2º-There's looseness on the handpiece coupling
3º-The air valve is not completely open
4º-There's insufficient air pressure for the equipment
5º-The air compressor is not working properly
6º-There's a blocking in the pneumatic system
1º-There's looseness on the handpiece coupling
2º-The gasket is worn
3º-The gasket is not properly sealing the coupling
4º-The coupling is worn
1º-The coupling spray's adjustment is closed
2º-The water tank is empty
3º-There's looseness on the handpiece coupling
4º-The air valve is not completely open
5º-There's insufficient air pressure for the equipment
6º-The hydro-pneumatic system is locked
1º-The saliva ejector's filter is clogged
2º-There's insufficient air pressure for the equipment
3º-The equipment's drain hose is obstructed
4º-The sewer is clogged
5º-The hydro-pneumatic system is locked
1º-There's a problem with the power supply
38
Solutions
1º-Connect the equipment to the mains electricity
2º-Switch On the electrical circuit breaker
3º-Call the power supply company
4°-Disable the Emergency Stop button
5º-Contact Olsen certified assistance
1º-Check troubleshooting item 1
2º-Repeat the recording, see the operation guide
3º-Contact Olsen certified assistance
1º-Call the air compressor assistance
2°-Contact Olsen certified assistance
1º-Check troubleshooting item 1
2º-Contact Olsen certified assistance
3º-Contact Olsen certified assistance
1º-Lubricate the handpiece ducts.
2º-Attach the handpiece correctly.
3º-Open completely the air valve.
4º-Check the inlet air pressure.
5º-Call the air compressor assistance.
6º-Contact Olsen certified assistance.
1º-Attach the handpiece correctly
2º-Replace the gasket
3º-Apply the original handpiece seal
4º-Contact Olsen certified assistance
1º-Align the coupling's green dots
2º-Fill the water tank
3º-Attach the handpiece correctly
4º-Open the air valve
5º-Open the main air valve
6º-Contact Olsen certified assistance
1º-Clean the saliva ejector's filter
2º-Open the main air valve
3º-Check if the sewer's drain pipes are crushed and
release the flow
4º-Check if there's water outflow on the spittoon
bowl and clean the sewer if necessary
5º-Contact Olsen certified assistance
1º-Check troubleshooting item 1