Primo Star / Primo Star iLED
3.1.3
Connecting the battery supply unit
Inserting batteries into the battery supply unit or remove them as follows:
• Loosen the four slotted screws of the battery supply unit.
• Remove the lid upward.
• Insert five commercial mono-cell batteries (D), type NiCd or NiMH, 1.2 V, with a capacity of 5000 mAh
(min.) to 9000 mAh (max.), paying attention to correct polarity (see markings in the battery
compartments).
Batteries of different types or different capacities must not be used together. Insert
rechargeable batteries only.
• Push the changeover switch for the battery type (Fig. 3/1) to the correct position:
ON = NiMH; OFF = NiCd
• Put the lid on again.
• Fasten it with the four slotted screws.
Primo Star iLED with reflected-light fluorescence illuminator:
• Pull the connector (Fig. 13/5) of the plug-in power unit (Fig. 13/6) out of the connection socket
(Fig. 13/1) of the reflected-light illuminator and insert it into the connection socket of the battery
supply unit (Fig. 13/4). The connector of the reflected-light illuminator (Fig. 13/3) is already connected
to the connection socket of the microscope (Fig. 13/7).
• Insert the connector (Fig. 13/2) of the battery supply unit into the connection socket (Fig. 13/1) of the
reflected-light illuminator.
• Insert the plug-in power unit (Fig. 13/6) into a power outlet.
Fig. 13
Connecting the battery supply unit
M60-2-0011 v 08/2008
Start-Up and Operation
Carl Zeiss
19