Maintenance of the Clean Head System
Outlet Holes
Outlet Holes
Fig. 4
8
To maintain effective performance of the NSK Clean Head System it
is important before steriliz ation to clean the specific outlet holes
located in the handpiece head. (Fig. 4)
Remove any debris in and around the outlet holes with the brush.
Half fill a cup with clean water. Operate the handpiece and, while it is
Fig. 5
running, immerse half of the handpiece head in the water, (Fig. 5)
Intermittently stop and start the handpiece in the water for about 5
seconds. While still running, remove the handpiece from the water.
Stop the handpiece and wipe it dry. If debris remains, repeat the
process until the area is clean.
Notice
Use only clean water to maintain the Clean Head System.
9