Prodigy
Important Control Solution Test Information
• Use only Prodigy
• Check the expiration date on the control solution bottle.
Do not use if expired.
• Control Solution, meter and test strips should come to
room temperature (68–77°F / 20–25°C) before testing.
• Use within a period of 90 days from the date that you first
open it. Record the discard date on the control solution
bottle and discard after 90 days.
• Store the control solution tightly closed at temperatures
below 86°F (30°C). Do not refrigerate.
Important: Prodigy
located on the back of the test strip vial. They are not
recommended target ranges for your blood glucose level.
Prodigy Diabetes Care, LLC
Talking Meter
Control Solutions.
®
meter's Control Solution ranges are
®
20
Owner's Manual
www.prodigymeter.com