STEP 3: Apply Control Solution
• Shake the control solution bottle
well then remove cap.
• Squeeze the bottle and discard
the first drop then wipe off the
dispenser tip with clean tissue
paper or cotton.
• Squeeze the bottle again to get a
second drop onto a clean,
non-absorbent surface or on
your fingertip. See Figure 19.
• Apply the drop to the opening of the absorbent channel
of the test strip (where it meets the narrow channel)
until the confirmation window is filled. See Figure 20
• The meter begins to count down. See Figures 19, 20
and 21.
Caution: To avoid contaminating the control solution with
the content of the test strip, DO NOT DIRECTLY APPLY
CONTROL SOLUTION ONTO THE TEST STRIP.
Prodigy Diabetes Care, LLC
Talking Meter
Figure 20
Figure 21
22
Owner's Manual
Figure 19
www.prodigymeter.com