Data Security; Guidance And Manufacturer's Declaration - Medtronic Guardian Connect Manual De Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 101

Data security

The transmitter is designed to only accept radio frequency (RF) communications from
recognized and linked devices. You must program the app to accept information from
a specific transmitter. Transmitted sensitive data is encrypted to prevent
unauthorized receipt or communication.

Guidance and manufacturer's declaration

Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The transmitter is intended for use in the electromagnetic environment specified be-
low. The customer or the user of the transmitter should make sure that it is used in
such an environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
RF emissions
Group 1
 CISPR 11
RF emissions
Class B
 CISPR 11
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The transmitter is intended for use in the electromagnetic environment specified be-
low. The customer or the user of the transmitter should assure that it is used in such
an environment.
Immunity Test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transi-
ent/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
The transmitter must emit electromagnetic energy in order to per-
form its intended function. Nearby electronic equipment may be
affected.
The transmitter is suitable for use in all establishments, including
domestic and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic
purposes.
IEC 60601 Test Level
±2 kV, ±4 kV, ±8 kV Air
±2 kV, ±4 kV, ±6 kV Indirect
±2 kV for power supply
lines
±1 kV for input/output lines
±1 kV line(s) to line(s)
±2 kV line(s) to earth
-23-
Compliance
Electromagnetic Envi-
Level
ronment - Guidance
±2 kV, ±4 kV,
For use in a typical do-
±8 kV Air
mestic, commercial, or
hospital environment.
±2 kV, ±4 kV,
±6 kV Indirect
Not applicable
Requirement does not ap-
ply to this battery pow-
ered device.
Not applicable
Requirement does not ap-
ply to this battery pow-
ered device.

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido