Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The transmitter is intended for use in the electromagnetic environment specified be-
low. The customer or user of the transmitter should assure that it is used in such an
electromagnetic environment.
Immunity Test
Radiated RF
 IEC 61000-4-3
Note:
At 80 MHz and 800 MHz, the higher frequency range applies.
Note:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption, and reflection from structures, objects and people.
IEC 60601 Level
Compliance Lev-
el
3 V/m
10 V/m
 80 MHz to
 80 MHz to
6 GHz
6 GHz
-25-
Electromagnetic Environ-
ment Guidance
Portable and mobile RF com-
munications equipment should
be used no closer to any part of
the transmitter, including ca-
bles, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Refer to the recommended sep-
aration distance table for more
information.
d = 0.35 P
 80 MHz to 800 MHz
d = 0.70 P
 800 MHz to 6 GHz
Where P is the maximum out-
put power rating of the transmit-
ter in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site sur-
a
vey
, should be less than the
compliance level in each fre-
b
quency range
.
Interference may occur in the
vicinity of equipment marked
with the following symbol: