4. Attach the protective cap.
5. Fit the cable clamp over the connecting cable and
secure with a bolt and washer at the tapped hole.
6. Connect the connecting cable to the measuring
chamber.
$
Andros measuring chamber connection (posi-
tion J9)
$
Siemens measuring chamber connection (posi-
tion X2)
Robert Bosch GmbH
Installation BEA 050 | 1 687 001 504 | 9
3.3
NO sensor tubing connection
i
The NO sensor is connected in series with the O2
sensor. Make sure the tubing is not kinked.
1. Remove the hose from the test gas outlet connection
and connect to the free connection of the NO sensor
receptacle.
2. Connect the hose of the NO sensor to the test gas
outlet.
3. Fit the housing cover.
3.4
NO sensor adjustment
i
Adjustment and calibration of the NO sensor are
described in the Online Help of the customer service
software "BEA service program" (Bea_kd.exe).
Adjust and calibrate the NO sensor as described in
the Online Help.
en
|
1 689 978 394
2016-08-08