Changing the fuses
Storage
Cleaning
Metrology checks
Repairs under
warranty and
outside of the
warranty
Repairs outside of
Metropolitan France
20
Open the device. Use the same type of fuses to ensure the safety of users
and of the device.
A LED comes on when one of the 1.6A and/or 10A fuses has blown and there
is voltage of VAC>110V between the terminals.
A and COM = (HS 1.6 and/or 10A fuse(s)),
10A and COM = (HS 10A fuse).
Also, for a blown 10A fuse:
between the VΩ and COM terminals, on the VDC calibres under 50VDC and
150VAC with voltage of 110VAC;
•
between the VΩ and COM terminals, on the VDC calibres under
500VDC and 1500VAC with voltage of VAC >220V.
If the multimeter is not used for a period of over 60 days, remove the battery
and store it separately.
For a shorter period, avoid leaving the multimeter in ohmmeter position.
There is a risk of premature battery wear if the tips come into contact with it.
•
The multimeter must be disconnected from all electricity sources.
•
To clean the casing, use a cloth dampened with soapy water. Wipe
over with a damp cloth. Dry quickly with a dry cloth or forced air.
Regular checks must be carried out as for all measurement or test
devices.
For checks and calibration, contact our COFRAC-approved metrology
laboratories or branches of MANUMESURE.
Information and contact details on request:
Tél. : 02 31 64 51 43
Send your devices to a regional Manumesure branch approved by Chauvin-
Arnoux.
Information and contact details on request:
Tél. : 02 31 64 51 43 Fax : 02 31 64 51 09
For all repairs under the warranty or outside of the warranty, return the device
to your retailer.
Fax : 02 31 64 51 09
English
Multimeter MX1