Descargar Imprimir esta página

Hach 50 11 Manual Del Usuario página 11

Publicidad

Idiomas disponibles

Idiomas disponibles

Life of an electrode
The average life expectancy of a pH electrode is approximately
one year depending on the working conditions.
An electrode used for more than six months that does not re-
sponse appropriately after regeneration, must be replaced by a
new one.
If an electrode has too short life span, less than three months,
ask us about it. Probably there is a more suitable electrode for
the application.
Specifications
pH measuring range
Operating temp.
Reference element
Diaphragm
Electrolyte
Body material
Min. immersion depth
Applications
Food industry, viscous media, dirty samples, etc.
Limitations
Samples with pH < 2.
Temperatures exceeding 80 °C.
Samples with low conductivity, distilled water.
Warranty
HACH guarantees this electrode against manufacturing defects.
Validity: 6 months from shipment date.
Limitations:
• Accidental damage.
• Inadequate applications.
• Non-fulfilment of the recommendations.
• Normal wear and tear of the electrode.
QUALITY CERTIFICATE
The attached electrode 50 11 has passed the quality test and
follows the next specifications:
• Asymmetry potential < ± 15 mV.
• Sensitivity, pH 4...7 (at 25 ˚C) > 98%.
• Response time, pH 4...7 < 20 s.
2 ... 14
0 ... 80 °C
sleeved Ag wire coated with AgCl
open, circular
solid
glass
15 mm
11

Publicidad