Supplemental Study
A total of 10 subjects were enrolled in a prospective,
supplemental study using a within-subjects repeated-
measures design where each subject served as his/her
own control. Five subjects used the M Acoustic Earhook,
while the remaining 5 used the M T-Mic and electrical
stimulation only. The data presented below are pooled
between the two cohorts.
Sentences in Quiet at 65 dBA
Across all subjects, similar sentence recognition scores
in quiet were observed between AutoSense OS and the
AutoSense OS Off program.  See the table below.
N
Mean (SD)
Median
Min, Max
Sentences in Noise at 65 dBA, + 5 dB SNR
Multi-Talker Babble
Across all subjects, better sentence recognition scores
in noise were observed when using AutoSense OS as
compared to the AutoSense OS Off program. See table
below.
2 l
a
S
iSt
uto
eNSe
a
S
Verage
Core
10
88.92 (11.068)
94.65
63.2, 96.7
oS
2 l
iSt
(%)
oS o
a
Verage
10
89.76 (11.677)
92.88
61.2, 99.3
a
S
uto
eNSe
ff
S
(%)
Core
41