ONDEMAND* DEVICE REFILL CHART
5
2.5
Re ll Time (minutes)
DELIVERY ACCURACY: When filled to the labeled volume, basal
flow rate accuracy is ±15% and bolus dose accuracy is +10/-20%
of the labeled rates when infusion is started 0-8 hours after fill and
delivering normal saline as the diluent at 22°C/72°F.
CAUTION: The ONDEMAND* device should be worn
outside clothing and kept at room temperature. Temperature will
affect solution viscosity, resulting in faster or slower flow rate.
ONDEMAND* device has been calibrated using Normal Saline (NS)
as the diluent and room temperature (22°C, 72°F) as the operating
environment. Flow rate will increase approximately 1.4% per
1°F/0.6°C increase in temperature and will decrease approximately
1.4% per 1°F/0.6°C decrease in temperature.
6
30
60
Rx only
= CAUTION: Federal (U.S.A.) law restricts this device to
sale by or on the order of a physician.
The product is covered by one or more of the following patents
6,936,035; 6,981,967.
Additional U.S. and Foreign Patents may be issued and/or pending.
*Registered Trademark or Trademark of Kimberly-Clark Worldwide,
Inc. or its affiliates. © 2012 KCWW. All Rights Reserved.
For more information, please call
(English only) or visit www.myON-Q.com for the latest
product information and Technical Bulletins.