R82 Kudu Guia Del Usuario página 41

Ocultar thumbs Ver también para Kudu:
Tabla de contenido

Publicidad

Service information
Maintenance of a medical device is wholly the responsibility of the owner of that device. Failure to
maintain a device in accordance with the instructions may invalidate the device's warranty. Furthermore,
failure to maintain a device may compromise the clinical condition or safety of users and/or their
caregivers.
1st Day
• Read the User Guide carefully before
use and store it in a safe place for future
reference
• Store the enclosed tools on/at the product
Daily
• Use a dry cloth to clean the product
• Visual inspection. Repair or replace any
damaged or worn-out parts
• Ensure that all fasteners are in place and
fitted correctly
• Check all fittings and buckles for signs of
worn out parts
• Be aware of any MAX mark indications
before adjustment
Weekly
• Check that all wheels are moving freely
and that all wheel locks/brakes are easily
activated
• Wipe off any residue and dirt from the
product using a cloth with warm water and a
mild detergent/soap and dry before use
• Tire pressure
Monthly
• All nuts and bolts on this product must be
checked and tightened to avoid unnecessary
failures
• Oil the swing-away parts. We recommend
you use a professional lubricant system
Annually
• Inspect the frame for cracks or signs for
worn-out parts and carry out annual service.
Never use a product which appears to be
substandard or faulty
Washing
The cushion covers can be washed at 40°C
using a mild detergent in a washing machine.
Warewashing machine
This product can be washed at 60° using a
mild detergent for 15-20 minutes in a washing
machine designed for medical devices. Use the
machine's drying function to dry the product.
Remove all accessories before washing the
main product. Clean the accessories separately.
Please allow the product to dry before reuse.
Disinfection
The product without accessories and cushion
can be disinfected with a 70% disinfectant IPA
solution. It is recommended to wipe off any
residue and dirt from the product, using a cloth
with warm water and a mild detergent/soap
without chlorine and let it dry before disinfection.
support.R82.org
Service interval
A detailed inspection of the product must be
made every 12 months (every 6 months for heavy
users) and every time the product is reissued for
use. The inspection should be carried out by a
technically competent person who has been
trained in the use of the product.
The lifetime of this product in normal use is 5
years if all maintenance and servicing is carried
out in accordance with the manufacturer's
instructions and demonstrably recorded.
Service Check List:
Check and adjust the following functions as
per their intended use. Repair or replace any
damaged or worn-out parts. Please check the:
• Seat plane adjustment. The seat plane
is adjustable and stays in position after
adjustment.
• Wheels are moving freely and that all wheel
locks/brakes are easily activated. Adjust if
required
• Quick release wheels are removable/
replaceable
• Anti-tips can be fixed in working and resting
position
• Tire pressure
• Spoke guards is without cracks and secured
to the wheel with the clips mounted on each
guard
• Before storage: Always deactivate wheel
locks/brakes
• Nuts and bolts on the product are tightened
• Oil the swing-away parts
• Product for cracks or signs for worn-out
parts
• All labels on the product are intact
41
EN

Publicidad

Tabla de contenido
loading

Tabla de contenido