Warning
Equipment malfunction –
- Magnetic and electrical fields are
capable of interfering with the
proper performance of the device.
For this reason make sure that all
external devices operated in the
vicinity of the CPM device comply
with the relevant EMC require-
ments. X-ray equipment, MRI de-
vices, radio systems and cell
phones are possible sources of in-
terference as they may emit higher
levels of electromagnetic radiation.
Keep the CPM device away from
these devices and verify its per-
formance before use.
- Refer repair and maintenance to
authorized persons.
- Route all cables below the device
frame to either side, ensuring that
they cannot get caught in the mov-
ing parts during operation.
- Inspect ARTROMOT
age and loose connections at least
once a year. Damaged and worn
parts must immediately be replaced
with original spare parts by author-
ized staff.
Caution
Preventing chafing and pressure
sores – If your patient is adipose,
very tall or very short, be sure to
prevent chafing and pressure sores.
Place the leg concerned in a moder-
ate abductive position, if deemed ap-
propriate.
®
-SP3 for dam-
50
Caution
Equipment damage –
- Check that the voltage and fre-
quency ratings of your local power
line are those indicated on the
nameplate.
- The leg support element withstands
a maximum continuous load of
20 kg.
- Do not allow any objects (such as
blankets, cushions, or cables) to
get caught in the moving parts of
the CPM device.
- Do not expose the ARTROMOT
SP3 device to direct sunlight, be-
cause some of the components
may reach inadmissibly high tem-
peratures.
®
-