4. Device setup

Note: For a better understanding of
each step, please fold out pages 3
and 224.
4.1 Connecting the
device, performance
check
1. Connect the power cord to socket
(18) of the device and connect the
mains plug to a wall outlet with a
non-fused earthed wire (100 to 240
Volt, 50/60 Hz).
2. Turn the power switch (16) on.
Caution
Equipment damage Be sure to lift
the moving bracket for dorsal exten-
sion / plantar flexion (15) a little, relie-
ving the strain on it, before you either
press the locking pins for the height
adjustment together or loosen the
clamping lever for adjustment of the
inclination (5).
3. Release the locking device (19). To do
so, first relieve the strain on the car-
riage by lifting the moving bracket for
dorsal extension / plantar flexion (15)
a little. Then press the two black pins
of the locking device together simul-
taneously (Fig. A).
4. Set the carriage to an angle of ap-
prox. 60°, then release the pins of the
locking device. Check that the pins
engage in the perforation (20).
5. Again, hold the moving bracket for
dorsal extension / plantar flexion (15)
and release the clamping lever for ad-
justment of the inclination (5). Then
lower the moving bracket for dorsal
extension / plantar flexion (15) until it
forms an angle of 45° with the base.
Close the clamping lever to lock the
carriage in that position (Fig. B).
6. Follow these steps to set the carriage
to the home position
®
ARTROMOT
-SP3
without patient chip card
Press the MENU key on the program-
ming unit twice to access programming
level 2 (with each key press, you ad-
vance one level).
Press the "new patient"
meter key and select this function (a
check mark appears in the circle next to
the function).
Press the START key. The CPM device
automatically enters the home position.
®
ARTROMOT
-SP3
with patient chip card
- Initial adjustment for new patients
Insert the original patient chip card
(21) into the programming unit (22).
Press the MENU key on the program-
ming unit three times to access pro-
gramming level 3 (with each key
press, you advance one level).
Press the "new patient"
meter key and select this function
(a check mark appears in the circle
next to the function).
Press the START key. The CPM de-
vice automatically enters the home
position.
- Adjustment with programmed
chip card
Insert the original patient chip card
(21) into the programming unit (22).
Press the START key.
The CPM device automatically enters
the home position.
51