Before replacing fuses, turn off the
®
ARTROMOT
-SP3 and disconnect the
device from the power line.
Use an appropriate tool to remove the
fuse holder situated between the power
switch and the power connector (Fig. 1).
Replace the fuses and reinsert the fuse
holder (Fig. 2). Ensure that the fuse hol-
der properly locks into place.
Fig. 1
Fig. 2

6.3 Transport

The following operating steps must
be completed before transporting the
ARTROMOT
®
-SP3:
1. Activate the "transport setting"
function
in the menu (see
also 5.3) and start the ARTROMOT
SP3.
2. Remove the footplate (7) as well as
its bracket (11). To do so, simply pull
it from the square-end guide at
motor B. (Fig. G)
3. For transport of the device, motor B
must be mounted on the right side of
the moving bracket (15) (marked R).
Refer to section 6.4 Conversion, if
required.
4. Turn off the ARTROMOT
power switch.
5. Disconnect the power cord and the
programming unit.
6. Relieve the strain on the carriage by
lifting it a little at the moving bracket
for dorsal extension / plantar flexion
(15) and open the clamping lever
which adjusts the inclination of the
carriage (5). Select an angle where
the carriage is parallel with the adju-
sted height. Tighten the clamping
lever to lock the carriage in this posi-
tion.
7. Again lift the moving bracket for dor-
sal extension / plantar flexion (15) a
little to remove the strain. Then loo-
sen the locking device for the height
adjustment by simultaneously pres-
sing the two black pins of the lok-
king device together (19). Select a
height where the locking pins engage
in the top hole (marked: Transport).
8. Only use the original shipping box
for transporting the device. ORMED
GmbH & Co. KG cannot be held
liable for transport damage, if the
device was not shipped in its original
shipping box.
®
-
9. Pack the programming unit (22) in
the supplied box.
10. Place the footplate bracket (11) with
the footplate (7) and the power cord
in the extra box provided.
66
®