Manual Disinfection - Renfert SYMPRO Manual Del Usuario

Ocultar thumbs Ver también para SYMPRO:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 67
D1
D2
D3
D4
D5
D6

Manual disinfection:

Disinfection process – lid & seal
Open the cleaning beaker� Remove the silicone seal from
the lid� Thoroughly rinse both objects (cover and seal) under
running water� Wipe the entire lid (outer and inner surfaces
including the thread) and the silicone seal (outer and inner
surface) with a wipe disinfectant (*)�
(*) Recommendation:
CaviWipes, Metrex
Fill the cleaning beaker with disinfection liquid
Remove the cleaning beaker magnets (not applicable for first
time use)�
Fill the cleaning bowl (incl� cleaning pins) with an instrument
disinfection fluid for heavy organic contamination (*). For fill
quantity, see Application Table�
(*) Recommendation:
Sporicidin, Contec / Metricide OPA plus, Metrex
Disinfection process – Cleaning the pins & inside the
beaker
Replace the disinfected silicone seal back into the lid or in
the case of the SYMPRO mini-cup, back into the cup�
On the control panel, set a runtime of at least 10 minutes for
higher intensity (*) and start the disinfection process�
(*) These values apply to the disinfection fluids recommend-
ed in D2 and in the Application Table�
Other settings may be required for other types of liquid or
concentrations�
Disposal of disinfection liquid
After the end of the program, remove and open the cleaning
beaker, keep the cleaning pins with the help of the magnet,
on the bottom of the cleaning beaker, pour out the disinfec-
tant and dispose of according to the manufacturer's instruc-
tions�
Disinfection process – Cleaning the outside of the bea-
ker and & sound protection ring
Remove magnets� Remove the silicone sound-proof ring /
adaptor ring from the cleaning bowl� Wipe the surface of the
cleaning bowl (incl� thread and sound-proof ring / adapter
ring) with a standard wipe disinfectant (*)�
(*) Recommendation:
CaviWipes, Metrex
Disinfection process – tweezers / Plastic cup with lid
(gray), help:ex discolor f
After use, place the tweezers and vessel into an instrument
bath with commercially available highlevel instrument disin-
fection liquid (*) for a high degree of organic soling and im-
merse according to the manufacturer's instructions�
(*) Recommendation:
Sporicidin, Contec / Metricide OPA plus, Metrex