6. Settings
On seca 882 and seca pragma 883 scales, it is possible for the customer to change
certain default settings to suit his requirements.
Setting the damping system
The damping system setting changes the transient response of the scale and thus the
time taken for the alarm to sound or weight to be displayed. Slight damping and thus rapid
determination of weight are very important for the limit value function, for example.
– Press the Start key with no load on the
scale.
se(a
8. 8 . 8 . 8 . 8 .
,
secutively in the display. The scale is
then automatically set to zero and
ready for operation.
– Press the FUNC key and keep it de-
pressed for 2 seconds.
FIL and the current setting appear in
the display:
0 = slight damping
s rapid determination of weight
1 = moderate damping
s normal determination of weight
2 = severe damping
s lslow determination of weight
– Press an arrow key if you want to
change the value set.
– Press the FUNC key again to store the
value displayed. The scale is then back
in normal weighing mode.
– Step on the scale and stand still. Ob-
serve the effect of the altered damping.
– Adjust damping again as described
above if necessary.
– To switch off the scale, press the Start
key again. The settings are retained.