Choosing A Location; Power Connection (Power Supply) - Sartorius Secura 35 Instrucciones De Instalación Y Manejo

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 24
t
t
t
t
1
2
t
t
3
Inserting the Weighing Pan
t
t
t
t

3.3 Choosing a Location

Select the right setup location:
− Set up the device on a stable, even surface that is not exposed
to vibrations (e.g., a weighing stone).
− Maintain free access to the device at all times.
Choose a location that is not subject to the following negative
influences:
− Heat (heater or direct sunlight)
− Drafts from open windows, AC systems and doors
− Vibrations during weighing
− Heavy traffic areas (personnel)
− Excessive moisture
− Electromagnetic fields
− Extremely dry air
Acclimatization
Condensation from humidity can form on the surfaces of a
cold device when it is brought into a substantially warmer area.
You should therefore let a device that has been disconnected from
its power source acclimatize for approximately 2 hours before
reconnecting it to the power.
Inserting the side panels:
− Mount the guide carriages (1)
of the side panel by inserting
them into the guide rails on
the top.
− Insert the side panel at the
bottom behind the guide rod (2)
of the draft shield.
Make sure that:
− the guide carriage (1) is
properly and fully positioned on
the guide rail.
− the side panels are positioned
behind the guide rod (2).
− the panels can be moved
smoothly and freely.
Inserting the upper panel:
− Insert the top panel on the top
of the draft shield guide.
− Push the panel carefully
backwards.
− Gently lift the panel by its grip
in order to push it below the
guide rod (3) of the guide.
− Close the cover by folding it
downwards completely.
Insert the weighing pan into the
opening of the collection container.
Push the upper and side panels
forward to close them.

3.4 Power Connection (Power Supply)

AC Adaptor Assembly
Using the wrong power plug adaptors may cause
fatal electric shock and damage the equipment.
Never plug the power plug adaptor into the socket
when it is disconnected from the device (Danger of
electrical shock).
Use the right power plug adaptor for your mains power supply.
t
t
A AC adaptor versions:
Bag
Region / Country
a) transparent Europe / EU (except
United Kingdom)
b) yellow
United Kingdom
c) blue
USA, China and Japan
d) red
Australia
The packaging of AC adaptor module 6971790 with line adaptors
g), h), and i) is labeled with 6971791.
t
t
Removing / Replacing the AC Adaptor
t
t
Power Connection / Safety Precautions
− Only use original Sartorius AC adaptors. The AC adaptor has an IP
rating of IP40 in accordance with EN 60529 / IEC 60529.
− Make sure that the voltage rating printed on the manufacturer's
ID label is identical to that of your local mains voltage.
− If the stated supply voltage or the plug design of the AC adaptor
does not comply with your country's standard, please inform the
nearest Sartorius representative.
− The power connection must be made in accordance with the
regulations applicable in your country.
Installation and Operating Instructions Manual Secura
English
Bag
Region /
Country
e) black
India
f) turquoise
South Africa
g) green
Korea
h) pink
Brazil
i) white
Argentina
Push (1) and slide (2) the power
plug adaptor (A) required for your
power supply into the opening of
the AC adaptor module (B). When
doing this, the AC adaptor needs to
lock into position.
Unlock (1) the AC adaptor and then
detach (2) it.
, Quintix
®
®