4. Remove all weight from the platform.
5. Once the stable indicator is shown on the display,
press the ZERO key
start blinking.
6. Use the numeric keys to enter the calibration weight
you are using.
7.
Press the ZERO key
display.
LoAd
8. Place the calibration weight on the platform.
9. Once the stable indicator is shown on the display,
press the ZERO key
the display.
SUCCES
10. Scale will enter a self-test mode. Remove weight
before self-test mode is complete. After completing
the self-test, the scale will enter the weighing mode.
MULTI-POINT CALIBRATION (LINEAR)
Multi-point calibration uses four calibration points: zero,
1/3, 2/3 and full weight capacity of the scale. Refer to
the calibration weights charts below.
CALIBRATION WEIGHT (lb.)
Model #
1/3 WEIGHT
H-5819
2 lb.
H-5820
7 lb.
H-5821
15 lb.
H-5822
30 lb.
CALIBRATION WEIGHT (kg.)
Model #
1/3 WEIGHT
H-5819
1 kg.
H-5820
3 kg.
H-5821
6 kg.
H-5822
14 kg.
1. Turn on the scale and press the TARE key
PO CAL
the self-test.
PO CAL
PAGE 4 OF 33
OPERATION CONTINUED
. Digits on the display will
Zero
LoAd
.
will be shown on the
Zero
SUCCESS
.
will be shown on
Zero
2/3 WEIGHT
FULL CAPACITY
7 lb.
15 lb.
30 lb.
55 lb.
110 lb.
2/3 WEIGHT
FULL CAPACITY
3 kg.
6 kg.
12 kg.
14 kg.
25 kg.
25 kg.
50 kg.
Tare
will be shown on the display.
2. To select multi-point calibration, press the TARE key
P 1 LCA
P I LCA
3. Press the ZERO key
display.
a. If calibrating in lb., continue to step 4.
b. If calibrating in kg., press the TARE key
4. Press the ZERO key
display.
LoAd0
5. Remove all weight from the platform.
6. Once the stable indicator is shown on the display,
press the ZERO key
display.
LoAd1
7.
Place 1/3 of the scale capacity on the platform.
8. Once the stable indicator is shown on the display,
press the ZERO key
display.
LoAd2
11 lb.
9. Place 2/3 of the scale capacity on the platform.
27 lb.
10. Once the stable indicator is shown on the display,
55 lb.
press the ZERO key
display.
LoAd3
5 kg.
11. Place the full scale capacity on the platform.
12. Once the stable indicator is shown on the display,
press the ZERO key
the display.
SUCCES
during
13. Scale will enter a self-test mode. Remove weight before
self-test mode is complete. After completing the
self-test, the scale will enter the weighing mode.
will be shown on the display.
Lb
Zero
.
G
will be shown on the display.
LoAd0
Zero
.
LoAd1
.
Zero
LoAd2
.
Zero
LoAd3
.
Zero
SUCCESS
.
Zero
Tare
will be shown on the
Tare
.
will be shown on the
will be shown on the
will be shown on the
will be shown on the
will be shown on