INSTRUCTIONS FOR USE
b. Prior to using this device, please read the instructions for equipment set up and
use. Familiarize yourself with the product before application on a patient.
c. To prevent patient and/or user injury and/or equipment damage, examine the
device and surgical-table side rails for potential damage or wear prior to use. Do
not use the device if damage is visible, if parts are missing or if it does not function
as expected.
d. To prevent patient and/or user injury and/or equipment damage, verify the
device attaching clamps completely touch the table-side rails and are firmly in
place. Test the locking mechanism to ensure no movement when elevated or
pushed.
e. Whenever practical, a patient should be placed in the leg holders prior to
anesthesia. An alert patient can communicate normal range of motion limitations
and, to some extent, such patient's opposing muscles will guard against
abnormal joint moving and stretching.
f. The patient's toes may extend beyond the end of the stirrup boot.
g. The product must be used with compatible clamps.
h. check for patient contact pressure points and consult the physician before use.
i.
Warn the user against using non-approved padding.
j.
To secure the patient's legs in the boots prior to making adjustments.
CAUTION:
a. Keep patient and caregivers away from the area between the gas spring and
the main support tube. Possible pinch point hazard when in extreme high
lithotomy.
b. Do not exceed safe working load shown in the product specification table.
4.2 Product Specifications:
Mechanical Specifications
Product Dimensions
Material
Safe Working Load on the device
Overall Weight of Complete Device
Storage Specifications
Storage temperature
Storage Relative humidity range
Operating temperature
Operating Relative humidity range
Electrical Specifications
Not Applicable.
Document Number: D-720378
Version: C
38" x 21" x 12" (96.5cm x 53.3cm x 30.5cm)
Stainless Steel, Alloy Steel, Aluminum,
Engineering Polymers, Bronze
720 lbs (326 kg)
14.8 lbs. Each (6.7 kg)
-29ᵒ C to +60ᵒ C
15% to 85%
This device is intended to be used in a
controlled Operating Room environment.
Not Applicable.