MANUAL DE INSTRUCCIONES CODIGO 80241 REV D
12
6
J
.P. SELECTA s.a.
7 Installation
7.1 Location
Locate the autoclave near an appropriate power source for the machine.
According to the current legislation, the autoclave should be located in a
way that the safety valve discharge outlet DO NOT point to any people or
could not reach them in case of steam outlet.
Install the autoclave over a fl at, stable horizontal surface, suitable for the machine
weight (see chapter 3 «Technical Specifi cations»), leaving a free space of at least
20 cm around the equipment.
Fix a hose to the outlet (13).
WARNING
If you use the machine for quick depressurization, please fi x a pressure hose.
Fix the hose in such a way that cannot move by the effect of the pressure at the
end of the cycle.
Do not hold the hose with the hands. It could reach temperatures higher than
60ºC.
8 Operation
8.1 Preliminary operations for switching on
-
Check that the lateral valve (12) is not in the «Drainage» position.
-
Place the heater element cover tray at the bottom of the vessel.
-
Fill the sterilization chamber with water till the heater element cover rack.
-
Plug the autoclave to the mains.
-
Press the main switch on (16).
-
Place the material to be sterilized.
-
Close the lid.
8.2 Opening and closing the lid
CLOSING
-
To close, just press over the lid with one hand and turn the knob (15) with the
other hand till the «CLOSED» position.
15
OPENING
-
It is not possible to open the lid if the pressure is different to 0 bar.
-
To open, press the button (6) and turn the knob (15) with the other hand till
the «OPEN» position.
15
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.grupo-selecta.com
Junio 2012
(Sujetas a modifi caciones sin previo aviso)
Tel (34) 93 770 08 77 Fax (34) 93 770 23 62
Pag.: 20