Electrical installation
Check the electrical installation of the enclosure where the autoclave will be
installed.
For your safety, we advise you to feed the autoclave with a magnetothermic
and a differential
Size
80L
150L
Water inlet installation
In areas with "hard" water (conductivity above 100uS / cm) it is necessary to
use water decalcification equipments. It is recommended to work with soft
water, to extend the life of the equipment
The water inlet must have a pressure between 3 and 4 bar.
If it is necessary, install the following:
- A pressure group, if the pressure of the water network is less than 2 bar.
- A hand-operated device if the water network pressure is higher than 4 bar.
Installation of drain drain
Located at ground level or below.
Material to resist temperature of 100ºC.
(Gray PVC plumbing is suitable)
Water hardness classification table:
Water type
Soft water
Slightly hard water
Moderately hard water
Hard water
Very hard water
mg/l: milligrams per liter of calcium carbonate water (CaCO
*fH: French grade (10.0 mg CaCO
*dH: German grade (17.8 mg CaCO
*eH: English grade (14.3 mg CaCO
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
ENE/2020
(Sujetas a modificaciones sin previo aviso)
Energy
Power (W)
230VAC
4000
400III + N
7500
mg/l
≤17
≤60
≤120
≤180
>180
Magnetothermic
25A + N
16A III + N
*fH
*dH
*eH
≤1.7
≤0.95
≤1.19
≤6.0
≤3.35
≤4.20
≤12.0
≤6.70
≤8.39
≤18.0
≤10.05
≤12.59
>18.0
>10.05
>12.59
/l)
3
/l)
3
/l)
3