17 TECHNICAL SPECIFICATION

d
Gauge
Air Temperature
(Ta)
Surface
Temperature (Ts)
External K-Type
Probe (Te)
Relative
Humidity (RH)
Specific
Humidity (SH)
Gauge Memory
Gauge & LCD
Operating Range
Power Supply
Battery Life
Gauge
Dimensions
Gauge Weight
Can be used in accordance with: BS 7079-B4, IMO MSC.215(82) ,
IMO MSC.244(83) , ISO 8502-4, US Navy NSI 009-32,
US Navy PPI 63101-000
d
Do not expose the gauge to temperatures outside the gauge and LCD operating range.
e
Accuracy ±0.75°C below 10°C (±1.35°F below 50°F).
f
Accuracy ±2°C (±4°F) with K type probes supplied by Elcometer. Probes supplied by other manufacturers
may vary.
g
At 1m/s
h
The batteries supplied with the gauge have a maximum working temperature of 45°C (113°F). Prolonged
use of the gauge in excess of this temperature may require the use of alternative batteries.
i
For the IMO PSPC (International Marine Organisation, Performance Standard for Protective Coatings),
the relative humidity, surface temperature and dewpoint should be recorded. The Elcometer 319 can be
used for this.
www.elcometer.com
Temperature Range
-20
°C
to 80
(-4°F to 176°F)
-20°C to 80°C
(-4°F to 176°F)
-20°C to 80°C
(-4°F to 176°F)
-40°C to 200°C
(-40°F to 392°F)
0 to 100% RH
0 g/kg to 325 g/kg
(0 gr/lb to 2275 gr/lb)
Model S & T: 10 sets of readings
Model T: 25,000 sets of readings in up to 999 batches
2 x AA Batteries or via USB Cable
Manual Logging: Greater than 40 hours
Interval Logging: up to 400 hours
(1 reading every 10 minutes)
174 x 75 x 35mm (6.85 x 2.95 x 1.38")
300g (10.5oz)
i
Accuracy
±0.5°C
°C
(±1°F)
±0.5°C
(±1°F)
±0.5°C
(±1°F)
±0.5°C
(±1°F)
±3% RH
±8% SH
-20°C to 80°C
(-4°F to 176°F)
h
- including batteries
Resolution
±0.1°C
(±0.1°F)
e
±0.1°C
(±0.1°F)
±0.1°C
(±0.1°F)
f
±0.1°C
(±0.1°F)
g
0.1%
0.1 g/kg
(0.1 gr/lb)
(rolling memory)
(backlight off)
I