Storage
Additional
Information
Manufacturer
Contact
EN-10
Warning: Drying time depends on several
.
variables, including: altitude, humidity, type of
wrap, preconditioning, size of chamber, mass
of load, material of load, and placement in
chamber. Users must verify that drying time set
in their autoclave yields dry surgical equipment.
Warning: These instructions were developed
using the guidance from AAMI TIR 12, ISO 17665
and AAMI ST79 and Stryker recommends users
observe these standards.
Trays.with.non-absorbent.liners.(such.as.plastic.
or.silicone-fingered.organizing.mats).can.cause.
condensate.to.pool..To.preserve.sterility,.remove.
devices.from.standing.solution.
N/A
Stryker.Endoscopy
5900.Optical.Court
San.Jose,.CA.95138
+1.800.624.2244,.+1.408.754.2000
www.stryker.com