fIgure 1.0 Panel buTTons
1.1
CusTom
Programmable favorite setting.
Default CW 1.5 Watts
1.2
debrIde
Programmable favorite setting for
debridement. Default preset CW 0.4 Watts
1.3
PerIo
Programmable favorite setting for the
treatment of apthous ulcers and periodontal
pockets. Default preset Pulse 1.5 Watts
1.4
CuT
Programmable favorite setting for fast cutting.
Default preset CW 1.2 Watts
1.5
emergenCy sToP
output OFF
1.6
baTTery
Charge level indicator (Green, Amber,
Red) flashes when battery is charging
1.7
dIsPlay
Laser power level indicator
1.8
uP
Adjusts laser power up by 0.1 watts
1.9
doWn
Adjusts laser power down by 0.1 watts
1.10
ready
with LED Indicator – toggles between
Standby and Laser Ready Modes
1.11
Pulse
Mode with illumination. Toggles between
CW and Pulse modes
NOTE: system defaults to Continuous Wave mode
(not illuminated)
1.12
audIo
Mutes audio. Toggles between audio
On/Off
1.13
aImIng beam
brightness up and down
NOTE: Preset labels are not intended to imply diagnosis or
treatment of oral disease.
a inTRODUCTiOn
The SOL Laser unit is a dental soft-tissue laser surgical/
debridement device. The SOL Laser may be used for a
variety of soft-tissue, gingival modification, and sulcular
therapy procedures.
Safety is paramount when using any energy-based surgical
instrument and your office should implement a safety
program for the SOL Laser. If your office does not already
have a safety officer, one should be appointed to be
responsible for understanding proper use, safe operation,
and maintenance of the SOL Laser system. Their duties
should include training office personnel in all aspects of
system safety and management of the SOL Laser and
all accessories.
manufaCTurer's laser IdenTIfICaTIon,
ouTPuT and sTandards label
The SOL Laser carries the Manufacturer's identification
label affixed to the bottom panel of the Control Module.
2
2
Turns active laser
Adjusts blue aiming beam
MODEL: ASSY SOL LASER
ITEM #: 033989100
LOT #:
MANUFACTURE DATE:
MANUFACTURED BY: DENMAT HOLDINGS, LLC
LOMPOC, CA 93436 USA
LASER CLASSIFICATION (PER IEC/EN 60825-1:2007): Class 4
MAXIMUM LASER OUTPUT: 3W CW; 5W P
FREQUENCY IN PULSED MODE: Fixed, 10Hz
PULSE DURATION: 0.05 seconds
WORKING BEAM WAVELENGTH: 808 nm (+5 nm)
AIMING BEAM WAVELENGTH: 405 nm (+10 nm) CW
POWER INPUT: 30 Watts + 12 VDC supplied from
100-240 VAC 50-60 Hz Model FW7362M/12
COMPLIES WITH IEC 60601-1, IEC 60601-2-22;
IEC/EN 60825-1:2007; IEC 60601-1-2, CLASS B PER
EN 60601-1-2 AS WELL AS TO US FEDERAL
REGULATIONS FOR LASER PERFORMANCE
STANDARDS SET OUT IN 21CFR 1040.10 AND
1040.11 EXCEPT FOR DEVIATIONS PURSUANT
TO LASER NOTICE 50 DATED JULY 26, 2001.
m
ompanying Documents
Attention, consult Acc
b WARnings & CAUTiOns
Visible and Invisible Laser Radiation – Avoid eye or skin exposure to
direct or scattered radiation
Above Warning is affixed to front panel of SOL.
Symbol for: Laser aperture at the end of the handpiece
Laser aperture warning label affixed to system hand-piece cradle.
Warning: Laser Safety Eye Protection MUST BE WORN by the
Operator, Patient, Assistant, and anyone present when the laser is
activated. Eye Protection must conform to Specification DIN EN207
Annex II of the Directive 89/686/EEC with optical density in 800nm-
818nm of OD 4+ such as NoIR Laser Company filter model DII
Warning: Do not use in the presence of combustible or combustion
supporting gases
Warning: Do not use in the presence of supplemental therapeutic
oxygen supplies
Warning: Always test activate the device outside the mouth before
using on a patient.
Caution: US Federal Law restricts the use of this device to use by or on the order of a dentist
Caution: Use of controls or adjustments, or performance of procedures other than
those specified herein may result in hazardous radiation exposure.
m
001374700 02/13
CLASS 4 Laser Product