pH electrode 53 00
Introduction
The 53 00 is a pH electrode for general use.
Low maintenance electrode, as it does not
require monitoring of electrolyte level.
SAFETY NOTE
Before use with hazardous substanc-
es, find out about the relevant haz-
ards, protective measures, and what
to do in an emergency.
Electrode preparation
• Remove the storage protector
• Verify that there are not air bubbles
inside the membrane
removed by shaking the electrode as a
clinical thermometer.
Installation
In immersion, insertion, flotation hous-
ings, etc. with PG 13,5 thread.
See Housing instruction manual.
In some cases the electrode is installed,
with the help of sockets and adapters,
directly onto pipes.
Cable. They are connected to coaxial
cables with an AS9 connector.
Calibration
See pH-meter instruction manual.
Measurements. Recommendations
See pH-meter instruction manual.
Ensure that the electrode's specifications are
adapted to the process.
Sample must cover the diaphragm
Maintenance
• Deposits forming on the electrodes,
caused by continual exposure of the sen-
sors to a dirty or contaminated medium, is
one of the most frequent causes of prob-
lems when measuring pH. When possible
it's recommended to use a housing with
automatic sensor cleaning system.
• When an electrode gets contaminated,
the simple rinsing with distilled water is
not suffi cient. The most suitable cleaning
solution is one which acts in a more selec-
tive way over the contamination (look for
regeneration solutions).
• Re-hydration. A "dry" electrode recovers
by immersing it in diluted HCl solution
for several hours. It is recommended to
read the information supplied with the
electrode's regenerating solutions.
Troubleshooting
... the meter always shows the same pH
value.
Short circuit in the electrode or cable.
Substitute the cable or the electrode.
... shifting of pH values.
Reference system contamination.
Replace the electrode
... unstable reading.
Diaphragm blockage.
Clean, regenerate, replace the elec-
trode.