Troubleshooting Module Installation
1.
Power off the oscilloscope.
2.
Remove the module and examine the contacts for damage or
contamination. Clean contacts if necessary using a 75% isopropyl
alcohol solution.
3.
Reinstall the module into the oscilloscope.
4.
Power on the oscilloscope, and perform the Checking Module
Installation procedure. (See page 6.)
a.
b. If the Application Module Detected display does not list the
5.
Power off the oscilloscope and move the module to another slot.
6.
Power on the oscilloscope, and perform the Checking Module
Installation procedure. (See page 6.)
a.
b. If the oscilloscope does not list the module after you try the
MSO3000 and DPO3000 Series Application Module Installation
If the Application Module Detected display lists the application
module, cleaning the contacts fixed the problem.
application module, go to step 5.
If the Application Module Detected display lists the module,
you have a problem with one of the module slots. Contact your
Tektronix service center.
module in every slot, contact your Tektronix service center.
Checking Module Installation
7