Installation; Installation Overview - Hach TU5300 sc Manual Básico Del Usuario

Ocultar thumbs Ver también para TU5300 sc:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 53
Figure 2 Product components
1 TU5300 sc or TU5400 sc
2 Wall mount bracket (two tubing clips on bracket)
3 Tubing clips
4 Tubing clip screws, 2.2 x 6 mm
5 Mounting screws, 4 x 16 mm

Installation

Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
Installation guidelines
Make sure that there is a floor drain near the instrument. Examine the instrument daily for leaks.
This instrument is rated for an altitude of 3100 m (10,710 ft) maximum. Use of this instrument at an
altitude higher than 3100 m can slightly increase the potential for the electrical insulation to break
down, which can result in an electric shock hazard. The manufacturer recommends that users with
concerns contact technical support.

Installation overview

Figure 3
shows the installation overview with no accessories and the clearances necessary. Refer to
the expanded manual on the manufacturer's website for the system overview with all of the
accessories.
10 English
6 Cleaning lid screws and washers for hot water
applications
7 Vial replacement tool
8 Flow regulator
9 Service bracket
10 Desiccant cartridge
C A U T I O N
N O T I C E

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Tu5400 sc

Tabla de contenido