Specification
Measurement units
Range
Method detection limit
Response time
Accuracy
Linearity
Repeatability
Stray light
Resolution
Air bubble compensation
Sample requirements
Calibration options
Verification options
Verification (RFID or
®
Link2SC
)
Certifications
Warranty

General information

In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer's website.
2
1 mNTU = 0.001 NTU
3
For the best results, operate the instrument at a flow rate of 200 mL/min when the maximum
particle size is 20 µm. For larger particles (150 µm maximum), the best flow rate is 350 to
500 mL/min.
4 English
Details
TU5300 sc: NTU, FNU, TE/F, EBC or FTU; TU5400 sc: NTU, mNTU
TE/F, EBC, FTU or mFTU.
0 to 1000 NTU, FNU, TE/F and FTU; 0 to 250 EBC
TU5300 sc: 0.002 FNU at 25 °C (77 °F) ( < 0.025 FNU range); TU5400 sc:
0.0001 FNU at 25 °C (77 °F) ( < 0.025 FNU range)
T90 < 30 seconds at 100 mL/min
± 2% or ± 0.01 FNU (the larger value) from 0 to 40 FNU
± 10% of reading from 40 to 1000 FNU based on Formazin primary standard at
25 °C (77 °F)
Better than 1% for 0 to 40 NTU based on Formazin primary standard at 25 °C
(77 °F).
TU5300 sc: 0.002 FNU or 1% (the larger value) at 25 °C (77 °F) ( > 0.025 FNU
range); TU5400 sc: 0.0006 FNU or 1% (the larger value) at 25 °C (77 °F) ( >
0.025 FNU range)
< 0.01 FNU
0.0001 FNU (0.0001 to 0.9999/1.000 to 9.999/10.00 to 99.99/100.0 to 1000 FNU)
Default: TU5300 sc: 0.001 FNU and TU5400 sc: 0.0001 FNU
Physical, mathematical
Temperature: 2 to 60 °C (35.6 to 140 °F)
Conductivity: 3000 µS/cm maximum at 25 °C (77 °F)
3
Flow rate
: 100 to 1000 mL/min; optimal flow rate: 200 to 500 mL/min
Pressure: 6 bar (87 psi) maximum compared to air, 2 to 40 °C (35.6 to 104 °F)
sample; 3 bar (43.5 psi) maximum compared to air, 40 to 60 °C (104 to 140 °F)
sample
®
StablCal
or Formazin: 1-point calibration (20 FNU) for 0 to 40 FNU measurement
range, 2-point calibration (20 and 600 FNU) for 0 to 1000 FNU (full) measurement
range or 2- to 6-point custom calibration for a measurement range of 0 FNU to the
highest calibration point.
Glass verification rod (solid secondary standard) ≤ 0.1 NTU, StablCal or Formazin
Verification of the measurement value by comparison of the process and lab
measurements with RFID or Link2SC.
CE compliant; US FDA accession number: 1420492-xxx. This product complies with
IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50.
Australian RCM.
1 year (EU: 2 years)