Hach TU5300 sc Manual Básico Del Usuario página 25

Ocultar thumbs Ver también para TU5300 sc:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 53
Do not to touch or scratch the glass of the process vial. Contamination or scratches on the glass can cause
measurement errors.
Note: Make sure that no particles fall into the vial compartment.
1. Push menu.
2. Select SENSOR SETUP>DIAG/TEST>MAINTENANCE>VIAL REPLACEMENT.
3. Complete the steps that show on the controller display. The date the vial was replaced is
automatically saved after the last screen shows.
Refer to the illustrated steps that follow to replace the vial. To protect the new vial from
contamination, use the vial replacement tool to install the vial.
At illustrated step 3, put the process head on its side on a flat surface if a service bracket is not
installed near the instrument.
N O T I C E
English 25

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Tu5400 sc

Tabla de contenido