Nuvoair AIR NEXT Manual Del Usuario página 10

Ocultar thumbs Ver también para AIR NEXT:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 51
CAUTION: RISK OF
CROSSCONTAMINATION
EN
Each new individual that uses Air
Next shall use a new unused dispos-
able NuvoAir disposable turbine. If
the turbine is not replaced between
individuals there is a high risk of
infectious cross contamination.
AIR NEXT TESTS
PARAMETERS SUCH AS:
(Forced Expiratory Volume
FEV1
in 1 second) The volume of air (l)
exhaled in the first second during
a forced maximal expiratory effort
after a full inspiration.
(Forced Vital Capacity) The
FVC
total volume of air (l) exhaled during
a forced maximal expiratory effort
after a full inspiration.
(Ratio) The calculated
FEV1/FVC
result obtained by dividing FEV1
with FVC.
(Peak Expiratory Flow) The
PEF
maximum flow rate (l/s) during a
forced maximal expiratory effort
after a full inspiration.
The final results displayed in the ap-
plication are always the best values
obtained after performing
the desired number of tests includ-
ed in a session. The recommended
number of tests per session is 3,
however the user can perform up
to 8 tests.
Displayed test are compared to a
reference table (GLI) and displayed
as percent predictive value
(%pred.).
The reference values for predicted
normal lung function (FEV1, FVC,
FEV1/FVC, PEF) are based on ethnici-
ty, gender, age and height.
Interpretation of spirometric results
should be made by a physician or
allied health care
professional with suffcient training
in the performance and interpreta-
tion of spirometry.
10
Note: Home use of Air Next is
intended only for indicative purpose.
Interpretation of the test results and
diagnosis shall be done by a
healthcare professional.
MAINTENANCE
Handle Air Next device carefully.
Store it in clean and moisture free
conditions. Before use, always check
that the device is free from dust,
contamination or any particles.
Air Next does not require profes-
sional maintenance. If, however,
the outside of the device needs to
be cleaned make sure to only use a
damp cloth without any detergents.
Air Next device uses replaceable bat-
teries. NuvoAir Disposable Turbines
are disposable.

Publicidad

Tabla de contenido
loading

Tabla de contenido