Explanation of the safety signs and
symbols marked on the device:
Manufacturer's name and address.
CE mark: indicates that the device
is certified that it conforms to the
0598
requirements of the 93/42/EEC
medical device directive.
IP classification: indicates that the
IP32
device is protected against solid
objects over 2.5 mm entering as well
as faling drops of water, if the case is
disposed up to 15° from vertical.
Waste electrical and electronic
equipment: Dispose accordingly.
Do not dispose as unsorted
municipal waste.
Manufacturer's LOT batch/lot code.
Follow operating instructions:
Indicates the need for the user
to consult the instructions for
important information.
The symbol is used in accordance
with IEC 60601-1-2:2014 (ED. 4)
for products including radio trans-
mitters and in accordance with
EN 300 328.
Type BF applied part: Device that
has conductive contact or medium
to long term contact with the
patient in order to fulfill the
intended use.

OPERATING ENVIRONMENT

Air Next has been designed for use
in a doctor's office, in a hospital
setting or at home.
Air Next is not intended for use
in an operating theatre nor in the
presence of inflammable liquids
or detergents, nor in the presence
of inflammable anaesthetic gases
(oxygen or nitrogen). The device is
not designed to be used in direct
air currents (e.g. wind), sources of
heat or cold, direct sun rays or other
sources of light or energy, dust, sand
or any other chemical substances.
9
NUVOAIR
DISPOSABLE TURBINE
Air Next is designed to be used with
a NuvoAir disposable single use
turbine. The disposable turbine is in-
dicated for single use single patient
only. The integrity and functionality
is guaranteed by:
• Never holding the turbine under
a jet of water or air and never let
it come into contact with high
temperature fluids.
• Not allowing dust or foreign
bodies to enter the turbine
sensor, in order to avoid incorrect
functioning and possible damage.
The presence of any impurities
such as hair, sputum, threads etc.
within the body of the turbine
sensor may seriously compromise
the accuracy of the measurements.
• Avoiding environmental
contamination by cleaning waste
products, the user must adhere
to all relevant regulations.
EN