TROUBLESHOOTING

If any abnormality arises during use, please check the following points
POSSIBLE
SYMPTOMS
CAUSES
Batteries have
Unit does not turn
run down.
on when the START/
STOP button is
pushed.
Battery polarities
have been positioned
incorrectly.
EE measurement error
Cuff has been placed
symbol shown on dis-
incorrectly.
play or the blood pres-
sure value is displayed
excessively low
Did you talk or move
(or high).
during measurement?
Shaking of the arm with
the cuff on.
E1 error symbol shown
Air circuit abnormality.
on display
Cuff tube may not be
plugged into monitor
correctly.
E2 error symbol shown
Inflation pressure
on display
exceeding 300 mmHg.
E3 error symbol shown
Error determining
on display
measurement data.
Note: If the unit still does not work, contact HoMedics Consumer Relations.
Under no circumstance should you disassemble or attempt to repair the
unit by yourself. Contact information for HoMedics Consumer Relations
Department can be found on the warranty page.
CORRECTION
Replace them with
four new AA
alkaline batteries.
Re-insert the batteries
in the correct positions.
Wrap the cuff properly
so that it is positioned
correctly.
Measure again. Keep
arm steady during
measurement.
Check cuff connection.
Measure again.
Switch the unit off, then
measure again.
Measure again.
29

SPECIFICATIONS

Power Source:
Four 1.5V DC (AA) batteries or 6V DC AC adapter
Measurement Method: Oscillometric
Pressure: 40~280 millimeters mercury (mmHg)
Measurement Range:
Pulse: 40~199 beats/minute
Pressure: ±3 mmHg
Accuracy:
Pulse: ±5% of reading
Pressure Sensor:
Semi-conductor
Inflation:
Pump driven
Deflation:
Automatic Pressure release valve
Memory Capacity:
60 memories for each User (120 total)
1 minute after last button operation
Auto-shut-off:
Operation
Temperature: 50°F~104°F (10°C ~40°C)
Environment:
Humidity: 15% ~ 90% RH
Temperature: -4°F~158°F (-20°C~70°C)
Storage Environment:
Humidity: Less Than 90% RH
Weight:
0.86 lbs (390 g) (without batteries)
Standard size cuff: 9"–13" (23-33 cm)
Arm Circumference:
Large size cuff:13"–17" (33-43 cm)
4.5"(L) x 5.9"(W) x 3"(H)
Dimensions:
114mm(L) x 150mm(W) x 76mm(H)
Accessories:
6V DC AC adapter, arm cuffs with tube, pouch
Internally Powered equipment
Type BF Equipment
IPx0-No special protection against the
ingress of water
Not suitable for use in presence of flammable
anesthetic mixture with air, oxygen or
nitrous oxide.
Continuous operation with short-time loading.
Note: These specifications are subject to change without notice.