Troubleshooting - HoMedics BPW-1005-EU Manual De Instrucciones

Ocultar thumbs Ver también para BPW-1005-EU:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 47

11. TROUBLESHOOTING

n
PROBLEM
ity,
t, and
LCD Display
shows
abnormal
result
LCD shows low
battery symbol
LCD shows
"Er 0"
LCD shows
"Er 1"
LCD shows
"Er 2"
ure
LCD shows
ld
"Er 3"
LCD shows
or
"Er 4"
LCD shows
"Er 5"
LCD shows
"Er 6"
LCD shows
"Er 7"
LCD shows
"Er 8"
LCD shows
"Er A"
POSSIBLE CAUSE
The cuff position was not correct or it
was not properly tightened
Body posture was not correct during
testing
Speaking, arm or body movement, angry,
excited or nervous during testing
Irregular heartbeat (arrhythmia)
Low battery
Pressure system is unstable before
measurement
Fail to detect systolic pressure
Fail to detect diastolic pressure
Pneumatic system blocked or cuff is too
tight during inflation
Pneumatic system leakage or cuff is too
loose during inflation
Cuff pressure above 300mmHg
More than 3 minutes with cuff pressure
above 15 mmHg
EEPROM accessing error
Device parameter checking error
Pressure sensor parameter error
Apply the cuff correctly and try again
Review the "BODY POSTURE DURING
MEASUREMENT" sections of the
instructions and re-test.
Re-test when calm and without
speaking or moving during the test
It is inappropriate for peoplewith
serious arrhythmia to use this Electronic
Sphygmomanometer
Change the batteries
Don't move and try again.
Apply the cuff correctly and try again
Measure again after five minutes. If the
monitor is still abnormal, please contact
the local distributor or the factory.
11 11
SOLUTION

Publicidad

Tabla de contenido
loading

Tabla de contenido