stainless steel platform is packed separately.
1.
remove set screw on bottom of scale.
2. level the scale. Adjust the 4 feet and reference
bubble on top of scale to level.
3. set platform in the locating holes in the top cover.
CAUTION! Do not press down with excessive
force – this can damage the load cell.
4. Attach Ac adapter to the power jack on the bottom
of the scale.
ZErOInG tHE dIsPlAy
1.
Press ZErO/EntEr key to reset the display to 0 before
weighing or counting.
NOTE: Scale has an automatic re-zero to
handle minor drifting or material accumulation
on the weighing platform. If the display shows
a weight while the platform is empty, press the
ZERO/ENTER key to zero the scale.
tArE
Before weighing items, weigh the container so its weight
can be deducted from the accumulated item weight:
1.
Press ZErO/EntEr to re-zero the display.
2. Place the container to be used on the weighing
platform. the display will show its weight. Press the tArE
key. container weight will be stored as the tare value
and is subtracted from the display. display shows 0.
3. nEt indicator is on. When adding items to the
platform only the item weight will be displayed.
NOTE: Scale can be tared a second time to
show only the weight of the items added after
you press the TARE key again. Display will show
a negative (-) value if the container is removed.
ZERO indicator will also be on.
WEIGHInG A sAmPlE
to determine the weight of a sample:
1.
Place empty container on the platform.
2. Press tArE to subtract container weight.
3. Place sample in container. display shows weight of
sample less the weight of the container.
PAGE 3 OF 18
INSTALLATION
5. Power switch is located on the bottom right side of
the scale.
BOTTOM VIEW
OPERATION
PErcEnt WEIGHInG
scale allows the sample weight to be shown as 100%.
then any other weight placed on the scale displays as
a percentage of the original sample.
For example:
1. 350g is placed on the scale.
2. Press % Key.
3. display shows 100.00%.
4. remove the 350g weight, then place a 300g weight
on the platform.
5. display shows 85.71% (300g is 85.71% of 350g).
6. Press Func key to return the scale to weighing mode.
NOTE: Scale may jump by large amounts if
small weights are used to set the 100% level.
PArts cOuntInG
While the display is showing a weight, pressing the Func
key starts the parts counting function.
1.
Before starting, tare the weight of container used
and leave the empty container on the platform.
2. Place the number of samples on the scale. the
number should match the predetermined sample
sizes: 10, 20, 50, 100 or 200 pieces.
3. Press the Func key. display shows "P 10" (looking for
a sample size of 10 parts.)
4. to change the sample size: Press the tArE key –
display will cycle through 10, 20, 50, 100, 200 and
back to 10.
Level Feet
Power Switch
Set Screw
AC Adapter