Vega WEIGHTRAC 31 Instrucciones De Servicio página 80

Sensor radiométrico para la medición del flujo de masa
Ocultar thumbs Ver también para WEIGHTRAC 31:
Tabla de contenido

Publicidad

11 Anexo
Mapping of Process Value Status
Hex
Quality
0x00
Bad
0x01
Bad
0x0C
Bad
0x10
Bad
0x1C
Bad
0x20
Bad
0x44
Uncertain
0x48
Uncertain
0x4C
Uncertain
0x80
Good (NC)
Mapping of User Error Codes to FF Field Diagnostics
Priority
Description
31
Hardware failure
30
Memory failure
29
Software failure
28
Paramererization corrupt
27
Undefined 27
26
Undefined 26
25
Paramererization error
24
Conflict in MGC
23
Communication error in MGC
22
MGC slave reports error
21
Undefined 21
20
Undefined 20
80
Sub-Status
Non-specific
Configuration Error
Device Failure
Sensor Failure
Out of Service
Transducer in MAN
Last Usable Value
Substitute
Initial Value
Non-specific
User Error Codes
40, 41
34, 35, 36, 37
80
72
-
-
16, 17, 25, 52, 53, 57, 66, 117,
120
121, 122, 141
8
38
-
-
Condition
Unexpected error
It was tried to set a wrong unit with FF interface
On user error codes: 16, 17, 25, 52, 57, 66, 72,
117, 120
Five or more internal communications have failed
while "Process Data Update". Last usable value is
displayed
On user error codes: 8, 34, 35, 36, 37, 38, 73, 80,
86, 121, 122, 141
On user error codes: 40, 41, 53, 68, 123, 124, 125
Transducer block is in mode "Out of Service"
Channel is not assigned to an AIFB
Channel is not available in running application
Transducer block is in mode "Manual"
Three or more internal communications have fai-
led while "Process Data Update". Last usable value
is displayed
On user error codes: 29
After startup of device or channel assignment in
AIFB was changed. The channel unit may be unk-
nown until next "Process Data Update"
On user error codes: 13
No errors concerning to channel handling
On user error codes: 0, 33, 45, 71, 126, 127
WEIGHTRAC 31 • Foundation Fieldbus
NE-107 Status
FAILURE
FAILURE
FAILURE
FAILURE
-
-
FAILURE
FAILURE
FAILURE
FAILURE
-
-

Publicidad

Tabla de contenido
loading

Tabla de contenido