PM 620T-IS
Marking Details
Baseefa10ATEX0012X............................................................................. ATEX Certificate number
II 2 G ...............................................................................................Equipment group & category
Ex ib IIC T4 Gb (-10°C ≤ Ta ≤ +50°C)........................................Hazardous location markings
IECEx BAS 10.0004X ................................................................................ IECEx Certificate number
...........................................................................................................................................CE Mark
PM 620T-IS .....................................................................................................Specific apparatus type
SN *******........................................................................................................................... Serial number
DoM: MMM YYYY ........................................................... Date of manufacture, month and year
Requirements and Conditions
Special Conditions for Safe Use
The lower ambient temperature is limited to -10°C.
IECEx Approvals
For the IECEx certificate (IECEx BAS 10.0004X), visit the IECEx website at:
www.iecex.com
Electrical Parameters
PM 620T-IS
Ui = 7.9 V
Ci = 0.22 nF
Ii = 500 mA
Li = 0
Pi = 0.7 W
vi | DPI 620G-IS Safety and Quick Start Guide-English
© 2016 Baker Hughes, a GE company – All rights reserved.