15 ERROR MESSAGES (continued)
Error Message
01, 02 and 03
04, 05 and 06
07 and 08
09
10
11
12
13
14
15, 16 and 17
d
Or your local Elcometer supplier.
www.elcometer.com
TABLE 4 (continued)
Causes
High voltage probe handle
ADC error.
High voltage probe handle
DAC error.
High voltage probe handle
EEPROM error.
High voltage probe handle
CRC error.
High voltage probe handle
connecting cable (curly
cable) fault.
Current leakage.
Handle not compatible.
Handle data invalid.
Handle not recognised.
Handle switch presses not
recognised.
Action to Take
Remove high voltage probe handle
and ret. If error persists, contact
d
Elcometer .
Remove high voltage probe handle
and ret. If error persists, contact
d
Elcometer .
Remove high voltage probe handle
and ret. If error persists, contact
d
Elcometer .
Remove high voltage probe handle
and ret. If error persists, contact
d
Elcometer .
Return high voltage probe handle to
d
Elcometer .
Return to Elcometer for software
upgrade.
Remove high voltage probe handle
and ret. If error persists, contact
d
Elcometer .
Remove high voltage probe handle
and ret. If error persists, contact
d
Elcometer .
Remove high voltage probe handle
and ret. If error persists, contact
d
Elcometer .
Remove high voltage probe handle
and ret. If error persists, contact
d
Elcometer .
R
d
en-26