Note:
It is your responsibility to:
- Check calibration of the device every two years.
- Never knowingly use a defective device.
- Immediately replace parts that are broken, worn, missing, incomplete, damaged or
contaminated.
- Contact the nearest factory approved service center should repair or replacement
become necessary.
- Further, the user of the device bears sole responsibility for any malfunction that
results from improper use, faulty maintenance, improper repair, damage or alteration
by anyone other than Riester or authorized service personnel.

1.6 WARNINGS AND CONTRAINDICATIONS

There  is  a  danger  of  life-­threatening  electrical  shock.  
For  electrically  isolating  the  device  from  supply  mains  pull  the  plug  from  
the  socket  outlet.  
 
Unplug  the  instrument  before  cleaning!  
 
At  the  proposed  RBP-­100  location  the  power  plug  must  be  accessible.  
 
Do  not  use  this  device  on  pediatric  patients  under  3  years  old,  infants,  or  neonates.  
 
The  Riester  RBP-­100  is  not  intended  for  continuous  monitoring.  Do  not  leave  the  
 
device  unattended  while  taking  measurements  on  a  patient.  
 
Do  not  operate  the  Riester  RBP-­100  near  flammable  anesthetics  or  
 
volatile  vapors.  An  explosion  may  result.  
 
Do  not  use  the  device  if  it  has  failed  its  diagnostic  self  test  or  if  it  displays  a  greater  
 
than  zero  pressure  with  no  cuff  attached.  
 
Do  not  make  repairs  yourself.  Equipment  must  be  returned  to  Riester  or  
 
authorized  service  personnel  for  repairs.  Substitution  of  a  component  
different  from  that  supplied  may  result  in  measurement  error.  
 
The  Riester  RBP-­100  is  not  intended  for  patients  connected  to  a  
 
cardiopulmonary  bypass  machine.  
 
If  Luer  Lock  connectors  are  used  in  the  constructions  of  tubing,  there  is  a  possibility  
 
that  they  might  be  inadvertently  connected  to  intravascular  
fluid  systems,  allowing  air  to  be  pumped  into  a  blood  vessel.  
The  Riester  RBP-­100  must  be  charged  before  using  it  for  the  first  time.  
 
For  accurate  blood  pressure  measurements,  ensure  that  the  
circumference  of  the  arm  fits  within  the  range  markings  on  the  cuff.  
 
Only  use  such  accessories  as  are  recommended  for  use  with  this  device.  
Compressing  the  pneumatic  tubing  may  cause  system  errors.  
Prevent  water  or  other  fluids  from  entering  any  connectors  or  vents  on  
the  device.  Should  this  happen,  all  connectors  should  be  dried  with  warm  air.  Then  
check  the  calibration  of  the  device  and  operating  functions  before  reusing.  
 
If  the  Riester  RBP-­100  is  dropped  or  mishandled,  please  have  it  checked  by  a  
authorized  service  center  before  bringing  it  back  into  use.  
At  least  every  three  months,  inspect  cords  and  accessories  for  fraying  or  other  
mechanical  damage.  Replace  as  necessary.  
Check  the  calibration  of  your  Riester  RBP-­100  at  least  once  every  two  
RBP-100