   illumination  should  activate.    
 
 
9.3     Cleaning  the  cuff    
Carefully  remove  spots  on  the  cuff  with  a  damp  cloth  and  soapsuds.  
 
WARNING:  Do  not  wash  the  cuff  in  a  washing  machine  or  dishwasher!    
 
 
9.4     Accuracy  test    
9.4.1. VERIFICATION OF CALIBRATION (STATIC)
We  recommend  this  device  is  tested  for  accuracy  every  2  years  or  after  mechanical  impact  (e.g.  being  
dropped).  Please  contact  your  local  Riester  Service  to  arrange  the  test  (see  foreword).  
 
Monitoring  of  instruments  
Test equipment:
All  countries  except  for  Germany:  
The  respective  legal  provisions  apply  for  all  countries,  except  for  Germany.  The  reference  manometer,  
which  is  used  for  calibration,  must  be  traceable  to  national  and  international  measurement  standards.  

Equipment Required:

9.4.1.1
Calibrated reference manometer (1) or equivalent (pressure reference).
 It  is  not  allowed  to  make  changes  to  the  device!  
9.4.1.2
500ml volume (3).
 
9.4.1.3
Hand Inflation Bulb (4) with Bleed valve (pressure control).
9.4.1.   Verification  of  Calibration  (Static)  
 
9.4.1.4
Blood pressure device (2) to be tested.
Test  equipment:  
 
To perform Calibration Verification, proceed as follows:
Equipment  Required:  
9.4.1.1   Calibrated  reference  manometer  (1)  or  equivalent  (pressure  reference).  
9.4.1.6
Make the necessary connections using the materials listed in the Equipment
9.4.1.2   500ml  volume  (3).  
Required list. See Figure 1 for pneumatic connections.
9.4.1.3   Hand  Inflation  Bulb  (4)  with  Bleed  valve  (pressure  control).  
9.4.1.7
Enter the calibration verification mode at the RBP-100: The device must be
9.4.1.4   Blood  pressure  device  (2)  to  be  tested.  
 
switched off. Press the Start / Stop button and turn on the power switch at the
To  perform  Calibration  Verification,  proceed  as  follows:  
same time.
9.4.1.6   Make  the  necessary  connections  using  the  materials  listed  in  the  Equipment  
 
Required  list.  See  Figure  1  for  pneumatic  connections.  
9.4.1.7   Enter  the  calibration  verification  mode  at  the  RBP-­100: The  device  must  be  switched  off  
 
Press  the  Start  /  Stop  button  and  turn  on  the  power  switch  at  the  same  time.  
 
 
Figure1:  Pneumatic  connections  for  verifying  BP  calibration.
 
 
 
 
 