illumination should activate.
9.3 Cleaning the cuff
Carefully remove spots on the cuff with a damp cloth and soapsuds.
WARNING: Do not wash the cuff in a washing machine or dishwasher!
9.4 Accuracy test
9.4.1. VERIFICATION OF CALIBRATION (STATIC)
We recommend this device is tested for accuracy every 2 years or after mechanical impact (e.g. being
dropped). Please contact your local Riester Service to arrange the test (see foreword).
Monitoring of instruments
Test equipment:
All countries except for Germany:
The respective legal provisions apply for all countries, except for Germany. The reference manometer,
which is used for calibration, must be traceable to national and international measurement standards.
Equipment Required:
9.4.1.1
Calibrated reference manometer (1) or equivalent (pressure reference).
It is not allowed to make changes to the device!
9.4.1.2
500ml volume (3).
9.4.1.3
Hand Inflation Bulb (4) with Bleed valve (pressure control).
9.4.1. Verification of Calibration (Static)
9.4.1.4
Blood pressure device (2) to be tested.
Test equipment:
To perform Calibration Verification, proceed as follows:
Equipment Required:
9.4.1.1 Calibrated reference manometer (1) or equivalent (pressure reference).
9.4.1.6
Make the necessary connections using the materials listed in the Equipment
9.4.1.2 500ml volume (3).
Required list. See Figure 1 for pneumatic connections.
9.4.1.3 Hand Inflation Bulb (4) with Bleed valve (pressure control).
9.4.1.7
Enter the calibration verification mode at the RBP-100: The device must be
9.4.1.4 Blood pressure device (2) to be tested.
switched off. Press the Start / Stop button and turn on the power switch at the
To perform Calibration Verification, proceed as follows:
same time.
9.4.1.6 Make the necessary connections using the materials listed in the Equipment
Required list. See Figure 1 for pneumatic connections.
9.4.1.7 Enter the calibration verification mode at the RBP-100: The device must be switched off
Press the Start / Stop button and turn on the power switch at the same time.
Figure1: Pneumatic connections for verifying BP calibration.
Seite 16 von 23
RBP-100