9.2 DEVICE CARE
Cleaning and disinfection of medical products are meant to protect patients, users, and
third persons and lead to value retention of medical products. Due to product design and
materials used, there is no possibility to define the maximum limit of re-processing cycles.
Lifetime of medical products is determined by its function and gentle use. Before sending
back defective products for repair, the described re-processing cycles have to be applied
and followed.
Use a soft cloth and follow one of the methods listed for cleaning the exterior of the device:
9.2.1 Mild soap and water
9.2.2 Hydrogen perxodie solution (3% diluted with water)
9.2.3 Sodium hypochlorite solution (1 to 10 dilution of household chlorite bleach in water).
Note:
- To prove if all display segments and the display illumination unit are working correctly,
press and hold the "+" (5) or "-" (6) buttons. All display segments appear and the display
illumination should activate.
-
To prove if all display segments and the display illumination unit are working correctly, press
and hold the "+" (5) or "-" (6) buttons. All display segments appear and the display
illumination should activate.
9.3 CLEANING THE CUFF
9.3 Cleaning the cuff
Carefully remove spots on the cuff with a damp cloth and soapsuds.
Carefully remove spots on the cuff with a damp cloth and soapsuds.
9.4 Accuracy test
We recommend this device is tested for accuracy every 2 years or after mechanical impact (e.g. being
dropped). Please contact your local Riester Service to arrange the test (see foreword).
9.4 ACCURACY TEST
Monitoring of instruments
All countries except for Germany:
The respective legal provisions apply for all countries, except for Germany. The reference manometer,
We recommend that this device is tested for accuracy every 2 years or after mechanical
which is used for calibration, must be traceable to national and international measurement standards.
impact (e.g. being dropped). Please contact your local Riester Service to arrange the test.
It is not allowed to make changes to the device!
Monitoring of instruments
9.4.1. Verification of Calibration (Static)
All countries except for Germany:
The respective legal provisions apply for all countries, except for Germany. The reference
Test equipment:
manometer, which is used for calibration, must be traceable to national and international
measurement standards.
Equipment Required:
9.4.1.1 Calibrated reference manometer (1) or equivalent (pressure reference).
9.4.1.2 500ml volume (3).
It is not allowed to make changes to the device!
9.4.1.3 Hand Inflation Bulb (4) with Bleed valve (pressure control).
9.4.1.4 Blood pressure device (2) to be tested.
To perform Calibration Verification, proceed as follows:
9.4.1.6 Make the necessary connections using the materials listed in the Equipment
Required list. See Figure 1 for pneumatic connections.
RBP-100
9.4.1.7 Enter the calibration verification mode at the RBP-100: The device must be switched off
Press the Start / Stop button and turn on the power switch at the same time.
WARNING: Do not wash the cuff in a washing machine or dishwasher!