5 TAKING A READING
1
Hold the probe by its sleeve
2
To take a reading, bring the probe down onto the surface whilst holding it perpendicular
3
For subsequent readings, lift the probe off and then replace it onto the coated surface
ü DO
Hold the probe by the probe sleeve
•
•
Gently place the probe onto the surface
•
Allow the sleeve to make contact with the
surface - to improve accuracy
4
The display will dim if inactive for more than 15 seconds and will go 'black' if inactive for the
period defined in Menu/Setup/Screen Settings/Screen Timeout. Either press any key, or tap
the gauge to awaken it (Model B, S & T).
4
The gauge will switch off automatically after 5 minutes of inactivity
4
- - -
indicates reading outside range of probe
6 CALIBRATING THE GAUGE
1
Press the Cal softkey
2
For alternative calibration methods, select Cal/Cal Method
3
Select Calibrate and follow the on-screen instructions
4
When prompted place the probe on the centre of the foil(s)
4
Foils can be stacked
4
Not all calibration methods are available for all gauge types
see 7 Calibration Methods
5
û
DO NOT
Drag the probe over the coated surface
•
•
Bang the probe down hard onto the surface
•
Allow the probe to hover over the surface as
this could result in a false reading
90°