7.
Troubleshooting list
Malfunction
Device not working
Motor can't be activated/
regulated
A change in direction of
rotation via the foot con-
trol is not recognized by
the device
Motor stops and can't be
activated
Only for GA835:
No coolant
Note: The pump only runs
when the motor is active
Elliptical shaft runs nois-
ily, knocks or doesn't
work
Handpiece not working
Detection
Cause
"Power On" indicator light
Mains power cord not plugged in
8 is not illuminated.
Device not switched on
Defective mains power cord
Fuses blown
Device faulty
When the pedal of the
Foot control not connected
foot control is activated
the motor does not start or
cannot be regulated
Foot control faulty
Device faulty
Only the "right rotation"
A macro elliptical shaft is con-
indicator light 11 is illu-
nected
minated
Micro elliptical shaft without a
sensor sleeve
Elliptical shaft sensor 13 faulty
Foot control faulty
Motor not running despite
Overload protection interrupts the
requesting an increase in
motor
speed
Indicator light
Spray nozzle blocked
"Pump On" 3 is illumi-
Tube not properly inserted
nated, coolant pump is
running
Leaky tube
Indicator light
Device faulty
"Pump On" 3 is illumi-
nated, coolant pump is not
running
-
-
-
-
Remedy
Plug the mains power cord into the device socket
22and into the mains power socket, see Connecting
the power supply
Switch on the device with the "Power On" switch 9,
see Switching the device on/off
Replace the mains power cord
Replace fuses, see Fuse replacement
Have the product repaired by the Aesculap Technical
Service, see Technical Service
Insert the plug of the foot control into the control
cable socket (connection socket for foot control) 17
(back of the device), see Connecting/Disconnecting
the foot control
Have the product repaired by the Aesculap Technical
Service, see Technical Service
Have the product repaired by the Aesculap Technical
Service, see Technical Service
Only right rotation mode is possible, see Operating
principle
Provide a sensor sleeve GA183R for the elliptical shaft
or use an elliptical shaft with an elliptical shaft sen-
sor
Have the product repaired by the Aesculap Technical
Service, see Technical Service
Have the product repaired by the Aesculap Technical
Service, see Technical Service
After a cooling phase the motor is ready for operation
again
Change the spray nozzle
Insert the tube, see Tube set (for GA835 only)
Replace tube fittings
Have the product repaired by the Aesculap Technical
Service, see Technical Service
Refer to instructions for use of elliptical shaft
Refer to instructions for use of handpiece
13