Step-by-Step Instructions
6.5.3

Over-Instrumentation

A red segment and a warning signal indicate that the file has passed the
Apex. The cursor on the tooth icon shows "OVER".
6.5.4

Completion of the Measurements

Before unplugging the Measurement cable from the device receptacle,
disconnect the lip clip and the connection hook from the patient.
Move the file stopper to the selected reference point on the tooth.
Gently remove the file from the canal and measure the apical length
between the stopper and the file tip.
As a safety precaution in order to avoid over-instrumentation, it is
recommended to proceed as follows:
- stop file progression in the canal at the point where the ProPex
indicates 'APEX';
- place the file onto an endodontic ruler and measure the apical
length;
- subtract a minimum of 0.5 mm from the measured file length.
B EN PIIL DFU WEB / Rev.07 / 02-2018 (Old ZF 190292.EN)
®
II
17/48