Procedure For Lip Clip, Hook And Fork - Dentsply Sirona ProPex II Manual Del Usuario

Ocultar thumbs Ver también para ProPex II:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 147
Cleaning, Disinfection and sterilization
7.3

Procedure for Lip Clip, Hook and Fork

Operation
Pre-Disinfection or
1.
Decontamination
2.
Rinsing
B EN PIIL DFU WEB / Rev.07 / 02-2018 (Old ZF 190292.EN)
Operating mode
1. Pulp and dentine remnants
must be removed from the
accessories immediately.
2. After using the accessories on
the patient, place them directly
into a dish filled with a suitable
cleaning and disinfection
solution (e.g. CIDEZYME
®
ENZOL
Enzymatic Detergent
Solutions, Johnson & Johnson
Medical, 0.8% for between
1 minute and 2 hours) for
cleaning, pre-disinfection and
interim storage.
3. Wash the accessories under
flowing sterile, deionized water
or in a disinfection solution at
least three times for a minute
each time, in order to remove all
visible traces of contamination
and remnants.
Abundant rinsing (at least
1 min).
Warning
1. Do not allow remnants to dry
on. Clean no later than 2 hours.
For visible impurities that are
observed on instruments, a pre-
cleaning is recommended by
brushing them manually with soft
material.
Only use clean, soft brushes to
manually remove contamination
and remnants, or a clean, soft
cloth or wipe that is only used for
this purpose. Do not use metal
brushes or wire wool. Check that
no visible contamination or
remnants remain, and repeat the
®
,
pre-cleaning process if
necessary.
2. Make sure that the products
are fully immersed.
3. The disinfectant solution
should be aldehyde free (to
avoid blood impurities fixation),
suitable for disinfecting the
accessories, and compatible
with the accessories.
Note that disinfectant used for
pre-treatment is only for
personal protection and does not
obviate the need for disinfection
once cleaning has been
completed.
Pre-treatment should never be
omitted.
Use clean water.
If a pre-disinfectant solution
contains a corrosion inhibitor, it
is recommended to rinse the
instruments just before the
cleaning.
23/48

Publicidad

Tabla de contenido
loading

Tabla de contenido